4.2.1 Start and Stop in the Basic Mode

If the device temperature control is stopped, you can start the temperature control by pressing the
start button on the touch screen. After finishing the temperature control, you ca stop the control by
pressing the stop button on the touch screen.
• If you try to control the instrument when the doors is opened, it will not operate with a door-open sound.
Please close the door and restart it.
• If the door is opened while the device is operating, the fan and heater will stop operating, and a door-open
warning message will be shown without a notification sound. If the door is closed within the 'Door Open Delay
Alarm' time entered in System Settings > Alarms [4.4.4 Refer to Alarm], the previous operation continues. If
the door is not closed within the entered time, a warning sound is issued a pop-up will appear. When the
warning pop-up is displayed, the instrument will not work even when the door is closed. To restart the
instrument again, please close the door, check the warning sign, and restart the instrument.
52412Mechnicaloven_ver_1.0
Auto Tuning
Setting Timer
Door Open/ Close
Sound On/ Off
Screen lock On/Off
37