SONICflex cem
SONICflex cem tip no. 12 - 0.571.5431
SONICflex cem tip no. 12 A - 1.006.2021
SONICflex cem tip - 0.571.7142

Advertisement

loading
Need help?

Need help?

Do you have a question about the SONICflex cem and is the answer not in the manual?

Questions and answers

Summary of Contents for KaVo SONICflex cem

  • Page 1 SONICflex cem SONICflex cem tip no. 12 - 0.571.5431 SONICflex cem tip no. 12 A - 1.006.2021 SONICflex cem tip - 0.571.7142...
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 88400 Biberach 88400 Biberach Germany Germany Phone +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 3: Table Of Contents

    2.4 Qualification of personnel....................8 2.5 Protective equipment ....................8 3 Product description......................3.1 Intended use ........................ 9 3.2 SONICflex cem ......................9 3.3 Technical Specifications ....................9 3.4 Identification of tip type....................10 3.5 Transportation and storage conditions................10 4 Startup and shut down .....................
  • Page 4: User Instructions

    SONICflex is a registered trademark of KaVo Dental GmbH. All other trademarks are property of their respective owners. KaVo Original Factory Repair In the event of a repair, please ship your product to the KaVo Original Factory Repair using www.kavobox.com. KaVo Technical Service...
  • Page 5 SONICflex cem 1 User instructions Information on the packaging Material number Production batch LOT T Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code UDI symbol CE mark for medical devices Medical device, labelling of medical devices...
  • Page 6: Hazard Levels

    SONICflex cem 1 User instructions Hazard levels The warning and safety notes in this document must be observed to prevent personal injury and material damage. The warning notes are designated as shown below: HAZARD In cases which – if not prevented – directly lead to death or severe injury.
  • Page 7: Safety

    4 In addition, expose the tips to approx. 10 N (1 kg) mechanical load without function. 4 Only perform the check with the SONICflex cem tip (without the attachment screwed in). Worn tips can fracture or be contaminated and thus cause injury or infection.
  • Page 8: Accessories And Combination With Other Equipment

    Working with sharp or pointed products can lead to injuries or infections. 4 Use a glove or finger guard when you test, insert and remove the tips. KaVo recommends always working with a rubber dam and a dust extraction system.
  • Page 9: Product Description

    ▪ Intended for dental treatment only. All other types of use of or modifications to the product are not permitted and can be hazardous. The SONICflex cem tip is used in combination with the SONICflex to vibrate in- lays/ onlays.
  • Page 10: Identification Of Tip Type

    KaVo SONICflex 2003 KaVo SONICflex 2000 (Tips with long thread) Tips with "A" can be used for: all versions of KaVo SONICflex quick 2008 (Tips with short thread) 3.5 Transportation and storage conditions CAUTION Startup after refrigerated storage. Malfunction.
  • Page 11: Startup And Shut Down

    Dispose of and recycle the packaging appropriately in accordance with current packaging law, employing waste management / recycling companies. Comply with the comprehensive return system. KaVo has had its packaging licensed for this purpose. Please comply with the regional public waste-disposal system.
  • Page 12: Using The Cem Attachment

    4 Check that the cem attachment and the cem tip are firmly seated before treatment. 4 Screw the cem attachment onto the SONICflex cem tip clockwise by hand. 4 Check that the cem attachment is firmly seated on the SONICflex cem tip. 12 / 24...
  • Page 13: Removing The Cem Attachment

    SONICflex cem 4 Startup and shut down | 4.3 Removing the cem attachment 4.3 Removing the cem attachment 4 Unscrew the cem attachment from the SONICflex cem tip anticlockwise by hand. 4.4 Removing the SONICflex tip 4 Place the torque wrench on the SONICflex tip and unscrew it in counter- clockwise direction.
  • Page 14: Operation

    (1-3) without pressure. This prevents excessive heating of the cem attachment. To prevent the inlays or onlays from becoming lodged, place the SONICflex cem tip on several points of the inlay/onlay.
  • Page 15 Expose the com- posite joints according to the composite manufacturer's instructions, then pro- cess and polish with the corresponding finishing instruments. You will find detailed information at www.kavo.com. 15 / 24...
  • Page 16: Reprocessing Steps In Accordance With Iso 17664-1 / Iso 17664-2

    Malfunction and material damage. 4 Reprocess it in a washer disinfector only. KaVo recommends the following products based on the compatibility of the ma- terials. The microbiological efficacy must be ensured by the disinfectant manu- facturer and proven by an expert opinion.
  • Page 17: Manual Reprocessing

    SONICflex cem 6 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 6.2 Manual Reprocessing Consumables required: ▪ Cloths for wiping the medical device. 4 Spray the disinfectant on a cloth, then wipe down the medical device and al- low the disinfectant to act according to the instructions of the disinfectant manufacturer.
  • Page 18: Automated Drying

    NOTE Please comply with the instructions for use of the washer disinfector. 4 In order not to affect the KaVo medical device, make sure that the product is dry on the inside and outside after completion of the cycle. 6.4 Care products and systems - Servicing Not applicable.
  • Page 19: Sterilisation

    SONICflex cem 6 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 6.6 Sterilisation 6.6 Sterilisation Sterilisation in a steam steriliser (autoclave) in accordance with EN 13060 / EN ISO 17665-1 CAUTION Contact corrosion due to moisture. Damage to product.
  • Page 20: Other Products

    7 Other products 7 Other products Available from specialised dental dealers. Consumables Material summary Material number Torque wrench 1.000.4887 SONICflex cem attachment 0.571.7142 Cleaning adapter Material summary Material number Cleaning adapter for SONICflex tips, 1.009.7875 long thread Cleaning adapter for SONICflex tips, 1.009.7876...