To open the case for fuse replacement:
1. Remove the test leads from the Meter.
2. Remove the Meter from its holster.
3. Remove two screws from the case bottom.
4. Separate the case bottom from the case top.
5. Remove the fuse from its holder and replace it with an 11 A,
1000 V, FAST fuse having a minimum interrupt rating of
17 000 A. Use only Fluke PN 803293.
6. To re-assemble the Meter, first attach the case bottom to the
case top, then install the two screws. Finally, insert the Meter
into its holster.

Cleaning

W Caution
Do not use abrasives, isopropyl alcohol, or
solvents to clean the case or lens/window.
Wipe the case with a damp cloth and mild detergent. Dirt or
moisture in the terminals can affect readings.