Installation and Operation
#4: Sensor signal via buffered voltage output (Buffered OUT)
System module
All
All
42
Terminal
13 to 24
28 to 33
Example for channel AI1
Terminal 13: Signal (+)
l
Terminal 14: GND (-)
l
Terminal 15: open
l
Prepare the cable screen with the shield connection set (VIB
6.725-100).
Connect to a buffered voltage output via a coaxial cable with BNC plug.
Instructions
Signal at plus (+),
l
Screen at minus (-),
l
Leave the screen terminal (S) open.
l