Using The Optional Cooling - Thermo Scientific MaxQ Mini 4450 Manual

Hide thumbs Also See for MaxQ Mini 4450:
Table of Contents

Advertisement

Using the Optional Cooling Coil
Either tap water or other user-supplied media from a
refrigerated circulator flows through the submerged cool-
ing coil.
1.
Adjust setpoint to the desired temperature.
2.
Connect the coolant hoses to the cooling coil fit-
tings on the back of the unit. Start the coolant
flowing through the cooling coil; flow rate should
be at least 6-8 liters per minute. The coolant
temperature must be at least 15ºC less than the
desired chamber temperature. The coolant flow
and temperature must be constant.
3.
Allow the chamber temperature to stabilize at
the setpoint.
4.
Readjust the safety thermostat.
5.
Lift the cover and load platform. Close lid.
O
PERATION
25