Service & Maintenance
8.
.
8.1 General
In view of the significance of the accuracy of measuring results, the test device must be treated with
care. Dirt must be avoided at all time, especially at the high-pressure connection. Flushing and cleaning
of the test device must be made as described in para. 5.2. After measuring is completed, the test device
shell is stored in the suitcase. The high pressure regulator connection should be closed with the white
protective cap. Function checks should be made every month, and if erroneous measurements are
suspected. It is advised to do a full service check and recertify the BAT kit at a WatchGas factory trained
company every 12 months.
8.2 Disorders
▪
If the hand wheel of the test device cannot be loosened while it is connected to a compressed
gas cylinder or compressor, check if connection is under pressure. In this case close the cylinder
valve or compressor valve and release air from test device.
▪
Leaks at the connection to the cylinder or compressor could be due to a defect gasket. In this
case replace the gasket.
▪
If the high pressure regulator or the detector tube holder is defective, they must be replaced.
▪
In case of other disorders the device should be returned for repair to WatchGas or any
authorized service station.