Page 1
SST Dock Quick start guide SST Dock | QSG | V 1.0 | EN | 28-04-23...
Page 2
CAUTIONS & BEST PRACTICES WARNING IMPORTANT SAFETY INFORMATION: • If the SST Dock appears to be damaged or is not func- READ FIRST tioning in the correct way, then please get in touch with WatchGas or a local representative. •...
Ensure fresh air inlet filter is installed. • If potentially harmful gas are used for calibrations, users must read the MSDS of the gas connected. The SST Dock must be attached to a venting system or be used in a well-ventilated area.
Page 4
INTRODUCTION The SST Dock is an automated test station that allows Bump and Calibration testing and is com- patible with the full SST Range in one module. It has the capabilities to be connected via Wi-Fi, BLE or Cellular and allows for desk, wall and hard case mounted dependent on your application. With two inlets it allows for calibration of the standard four gases plus another gas type.
Page 5
Installing the brackets SWITCH ON THE SST DOCK If the SST Dock is fully discharged, please allow 5 min of charging prior to switching on the unit. LED will show red switching to white when prior to start-up. Connect the SST Dock to Mains by means of the wall adaptor included. The Status light turns on.
Watch Tab enabled bottles. Its important to fill in all data as the IMPORTANT: If optional pressure sensors are utilized these shall be activated or the SST Dock Dock produces a full calibration certificate. will ignore the signal Enter the password To add a new Calibration bottle click “Add bottle”...
Page 7
If the touch screen is in sleep mode – double open during operation. tap the screen to wake up the unit. The screen NOTE: The SST Dock will not perform any bump will instruct the next step. or calibration if a gas is not matching the detec- tor target gas.
Page 8
3-year SST1 then the device will already be activated. Turn on NFC. Step 2 Lift the lid on the SST Dock by using the two finger guides and the red latch. Step 3 Insert device by locating the alligator clip into the guide in the SST Dock.
Data Storage More than 1 year of operation Connection capability Wi-Fi, Cellular, BLE – Connected into the Cloud and WatchGas App Gas Inlets Fresh air inlet, 2 Gas inlets and one purge Bump Test, Calibration (optional), Device Configuration and Data Testing Function Download.
Best practices for maintenance check that inlet filter is free of dirt and replace if required. SST Dock can be cleaned with a soft, damp cloth. Do not use solvents, soaps, or polishes and do not immerse the SST Dock in liquids. Completely turn off unit and disconnect power supply before cleaning.