5.2.4 Mounting the sensor cleaner (/JC
option)
• After fixing the sensor with the cleaning
cage it is possible to screw the cleaner
tubing on the cage using the swage
lock coupling material provided with the
cleaner.
• It is recommended to fix the tubing to the
holder using tie wraps.
• Keep some flexible length of tubing for
ease of maintenance.
The cleaner can be used for both water and
air. It is recommended to use a solenoid
valve of the normally closed type with
matching connections. The solenoid valve
can be switched with the wash contact of
the EXA DO-converter.

5.3 Installing the floating ball fitting

5.3.1 Checklist

The floating ball fitting with sensor for
Dissolved Oxygen will be delivered in parts
and has to be assembled before it can
be used. Check if all standard parts are
delivered, see Figure 4. Please note that
items (4) and (2 + 6) are optional and, if not
specified with the order, these parts will not
be delivered. The position ring (9) will only
be received when PB30-00. is ordered. This
part is only needed when 11/2" pipe will be
used in stead of 50 mm.

5.3.2 Installation

Mount the floating ball not exactly vertical, but under a slight angle, so that air bubbles do
not 'stick' or adhere to the membrane.

5.3.3 Assembling the floating ball fitting (ref to Figure 4)

• Attach the mounting bracket (1) to wall or pipe; clamps are already fixed to the bracket.
• Install the main support arm (2) (in case of PB30-00... the customers pipe), in the
mounting bracket (1) and adjust its length to the desired length.
• Use the supplied adhesive (7) to stick the lower support to the main support arm (2).
When a 11/2" pipe is used, the position ring (10) must be mounted in between them to
compensate for the difference.
• Attach (not for PB30-00...) the 90° elbow (6) to the upper end of the main support arm
and cement it with adhesive (7).
• Guide the cable assembly (4) through the main support arm.
• Position the complete sensor (4) in the lower support.
• Place the lower support (3) into the ball (8) by screwing the fixing nut hand-tight.
• Connect the leads of the cable to the EXA DO converter.