A13.2602
Student Biological
Instruction Manual
.
To ensure the safety and obtain satisfactory performance, please study this instruction manual thoroughly bef
ore start to use the instrument.

Advertisement

loading

Summary of Contents for OPTO-EDU MDJ A13.2602

  • Page 1 A13.2602 Student Biological Instruction Manual To ensure the safety and obtain satisfactory performance, please study this instruction manual thoroughly bef ore start to use the instrument.
  • Page 2: Table Of Contents

    CONTENTS 1. Applications ................3 2. Specification ................. 3 3. Installation and Operation ............ 5 4. Maintenance ................ 7 5. Troubleshooting ..............9...
  • Page 3 Fig.1 MDJ200 Metallurgical Microscope Fig.2 MDJ100 Metallurgical Microscope...
  • Page 4: Applications

    1. Applications The MDJ series Metallurgical Microscope is widely used for identifications and analysis of the structures of different metals and alloys in factories,colleges and scientific department. 2. Specification 2.1. Total Magnification Objective Eyepiece 100X 250X 400X 100X 200X 500X 800X 12.5X 125X...
  • Page 5 2.3. Eyepieces Eyepiece Field Dia. Focus Eyepiece Type Magnification Mark (mm) (mm) WF10×-20 Plan Wide Field 10× φ20 Reticle Eyepiece WF10×-18 10× φ18 (0.1mm) WF16× Plan Eyepiece 16× φ13 15.6 WF12.5× Plan Eyepiece 12.5× φ14 5× Huygenian Eyepiece 5× φ20 2.4.
  • Page 6: Installation And Operation

    3. Installation and Operation 3.1. Installation 3.1.1 Put the body upright on the table steadily. 3.1.2 Put the head into body. 3.1.3 Put out the dust cover from eyepiece tube,and insert available eyepiece into eyepiece tube. 3.1.4 Turn around the coarse focusing knob to araise the stage.Remove the dust covers of nosepiece and put the objectives into the holes in nosepiece as per the power of objectives.
  • Page 7 same as the right eyepiece. 3.2.5 Field Iris Adjustment (As shown in Fig.1 and Fig.4) Turn the field iris ring to make the image of field iris smaller than the eyepiece field, and you can see the image of the field iris. Make the field iris center nearly overlap the eyepiece view center by adjusting the iris center screw.
  • Page 8: Maintenance

    3.2.8 Usage of Digital Camera 3.2.8.1 Get the clear image, then put out the eyepiece. 3.2.8.2 Connect the camera adaptor with camera, then insert the adaptor into the tube, and fix it with screw. 3.2.9 Usage of immersion oil objective If you must to use oil objective, it is necessary that adding...
  • Page 9 to microscope and worktable. 4.3 Replacement of bulb 4.3.1 Turn off power, and pull out plug. 4.3.2 Wait the bulb become cool. ▲ Please be sure that the bulb is cool, then follow by the nest operations. 4.3.3 Lay aside the microscope reliably, unscrew the knurled thumb screw of the lamp housing cover on the underside of base.
  • Page 10: Troubleshooting

    5. Troubleshooting In the period of using MDJ microscope, if there is any trouble occurs, please referring to the following sheet listed some common troubleshooting resolve them. Trouble Causation Remedy Plug is unreliable Plug in again Switch on but Bulb is bad Change bulb bulb dark Fuse is broken...
  • Page 11 Metallurgical Microscope...