Selecting »Mask« Application Mode
may be selected at time of start-up or during operation.
Evita 4
Switch ventilator to standby using the »
Touch » Tube/Mask « screen key.
Touch » Mask « screen key and confirm with dial knob, the
yellow "LED" in the screen key is now lit.
The mask symbol
is now showing in the status bar
instead of the tube symbol. The background color of the
status page will switch from blue to green.
Evita 2 dura
Switch ventilator to standby by pressing the »
for approximately 3 seconds.
Mute audible alarm with the » Alarm Reset « key.
Press » Tube/Mask « key.
The "Tube/Mask" menu will appear.
Turn dial knob to select » Mask « screen key.
Press dial knob to confirm.
NIV will appear in the status display.
The background color of the status page will switch from
blue to green.
« key.