Leak Compensation In "Mask" Application Mode - Dräger NIV Operating Instructions Addendum

Mask ventilation
Table of Contents

Advertisement

The following alarm messages are not displayed in »Mask«
application mode
— P. supp. > 4 s
— P. supp. > 1,5 s
— P. supp. > Tinsp
— Leakage
WARNING !
Always check alarm limits when switching from »Mask«
to »Tube«!
Leak Compensation in »Mask« Application
Mode
For the purpose of patient triggering, Evita 4 and Evita 2 dura
compensate leaks according to the selected patient mode up
to the following values:
Adult:
30 L/min
Pediatric:
15 L/min
Neonatal:
7 L/min
The ventilators compensate calculated leaks (independent
of patient mode) up to 200 % of the set tidal volume or a
maximum of 2 L.