Installing The Niv-Option; Using Niv - Dräger NIV Operating Instructions Addendum

Mask ventilation
Table of Contents

Advertisement

Installing the NIV-Option

NIV - Non-invasive Ventilation

Using NIV

Installing the NIV-Option
WARNING !
Installation of the NIV option should only be performed by
factory trained and authorized service personnel follo-
wing the respective installation documentation.
Using NIV
WARNING !
Deadspace will increase when using a mask.
Observe mask manufacturers data on dead space.
WARNING !
Switching to operation mode "mask" when ventilating an
intubated patient, could result in inadequate monitoring
and put intubated patients at risk.
WARNING !
Always check alarm limits after switching from "mask" to
"tube" mode.
WARNING !
When ventilating with a mask, avoid high airway pressu-
res.
Danger of aspiration.
In » Mask « application mode, all ventilation modes are available
with the exception of » ILV «.
NOTE: Automatic tube compensation (option A
have been activated during » Tube « application mode is
without effect in » Mask « application mode