  • Page 1 Operation Manual Shodex IC YS-50 Version 200701E Operation Manual Shodex™ IC YS-50 Sales Office: Technical Support: Showa Denko Europe GmbH info@shodex.de Shodex Business +49 (0)89 / 93 99 62 37 Konrad-Zuse-Platz 3 81829 Munich Quotes and Orders: Europe, Middle East, Africa, Russia Germany order@shodex.de...
  • Page 2 Certificate of Analysis (CoA) are matching and correct. Introduction Thank you for purchasing the Shodex™ product. Shodex IC YS-50 is an ion chromatography column suitable for simultaneous analysis of monovalent and divalent cations. It is applicable for both suppressor and non-suppressor methods. Column Components...
  • Page 3 Operation Manual Shodex IC YS-50 Version 200701E Usable Conditions Flow Rate (mL/min) Temperature Range (°C) Product Maximum Pressure pH Range Name (MPa) (per column) Recommended Maximum Recommended Range IC YS-50 2 - 12 20-60 IC YS-G 2 - 12 20-60 Usable solvents are listed below (1) The standard eluent is 4 mM methanesulfonic acid.
  • Page 4 Operation Manual Shodex IC YS-50 Version 200701E 6.2. Column Installation (1) Connect the column to LC system by following the “flow direction arrow” () indicated on the name tag. If guard column is used, position the guard column in front (before the inlet) of the analytical column.
  • Page 5 Operation Manual Shodex IC YS-50 Version 200701E Note • Use of on-line degasser is recommended. • It is recommended to set the pump limiter to avoid exceeding the maximum pressure. 6.4. Column Cleaning Problems in peak shapes and elution time changes or elevated column pressure etc. are often caused by insoluble or adsorbing components present in the eluent and reagents being deposited inside the column.
  • Page 6 Operation Manual Shodex IC YS-50 Version 200701E Column Inspection Inspection method is described in the Certificate of Analysis (CoA). Theoretical Plate Number (N) and Asymmetry Factor (Fas) were calculated using the below equations. (1) Theoretical Plate Number (N) (2) Asymmetry Factor (Fas) Additional Warnings (1) Do not remove end fittings.
  • Page 7 Version 200701E 10. Warranty (Version 4) (1) Showa Denko K.K. warrants that the Shodex™ product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.