Summary of Contents for Showa Denko Shodex IC SI-90 4E

  • Page 1 Operation Manual Shodex IC SI-90 4E Version 211013E Operation Manual Shodex™ IC SI-90 4E Sales Office: Technical Support: Showa Denko Europe GmbH info@shodex.de Shodex Business +49 (0)89 / 93 99 62 37 Konrad-Zuse-Platz 3 81829 Munich Quotes and Orders: Europe, Middle East, Africa, Russia Germany order@shodex.de...
  • Page 2 Operation Manual Shodex IC SI-90 4E Version 211013E Operation Manual Shodex IC SI-90 4E (Please read this manual carefully to achieve the best and consistent column performance for a long time) Important Handling Instructions Caution! • Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column and understand their proper handling methods to prevent potential health hazards or death from occurring.
  • Page 3 Operation Manual Shodex IC SI-90 4E Version 211013E Usable Conditions Maximum Temperature (°C) Flow Rate (mL/min) Product Pressure (MPa) Name Range Recommended Maximum Recommended Range (Per Column) IC SI-90 4E 3 - 12 20 - 60 IC SI-90G 3 - 12 20 - 60 Usable solvents are listed below.
  • Page 4 Operation Manual Shodex IC SI-90 4E Version 211013E (4) For the samples containing hydrophobic impurities, remove the impurities prior to the sample injection. The use of a solid phase extraction is recommended. Note • Use of guard column is recommended to protect the analytical column.
  • Page 5 Operation Manual Shodex IC SI-90 4E Version 211013E 7.3. Solvent Exchange When replacing the solvent, start the system at less than 0.5 mL/min. Recommended solvent volume to introduce at each step is 3 to 5 times of the column volume.
  • Page 6 Operation Manual Shodex IC SI-90 4E Version 211013E Attention! • Leaving the washing solution in the column for an extended time can damage the column. Do not suspend the column cleaning but complete all procedures until the last step to replace it with the eluent.
  • Page 7 3-3 If the Shodex product is disposed of 3-4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K.K. 3-5 If the performance of the Shodex product is not conform to the specification of the attached...