Summary of Contents for Showa Denko Shodex ODP2 HP-2B

  • Page 1 Operation Manual Shodex ODP2 HP-2B, 2D Version 211124E Operation Manual Shodex™ ODP2 HP-2B, 2D Sales Office: Technical Support: Showa Denko Europe GmbH info@shodex.de Shodex Business +49 (0)89 / 93 99 62 37 Konrad-Zuse-Platz 3 81829 Munich Quotes and Orders: Europe, Middle East, Africa, Russia Germany order@shodex.de...
  • Page 2 Operation Manual Shodex ODP2 HP-2B, 2D Version 211124E Operation Manual Shodex™ ODP2 HP-2B, 2D (Please read this manual carefully to achieve the best and consistent column performance for a long time) Important Handling Instructions Caution! • Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column and understand their proper handling methods to prevent potential health hazards or death from occurring.
  • Page 3 Operation Manual Shodex ODP2 HP-2B, 2D Version 211124E Column Specifications Product Column Size (I.D. Particle Pore Theoretical Plate Product Name Code x Length) (mm) Size (µm) Size (Å) Number (Per Column) F7622004 ODP2 HP-2B 2.0 x 50 ≥ 3,000 F7622005 ODP2 HP-2D 2.0 x 150...
  • Page 4 Operation Manual Shodex ODP2 HP-2B, 2D Version 211124E Eluent Preparation (1) Degas the eluent fully to prevent the formation of air bubbles. (2) Presence of small debris or insoluble substances may result in deterioration of the column and/or they appear as noise on the chromatograms. Filter the eluent with a 0.45 μm disposable filter to prevent the problems.
  • Page 5 Operation Manual Shodex ODP2 HP-2B, 2D Version 211124E 7.2. Column Installation (1) Connect the column to LC system by following the “flow direction arrow” () indicated on the name tag. If guard column is used, position the guard column in front (before the inlet) of the analytical column.
  • Page 6 Operation Manual Shodex ODP2 HP-2B, 2D Version 211124E 7.4. Column Cleaning Problems in peak shapes and elution time changes or elevated column pressure etc. are often caused by insoluble or adsorbing components present in the eluent and reagents being deposited inside the column.
  • Page 7 Operation Manual Shodex ODP2 HP-2B, 2D Version 211124E Column Storage Remove the column from the system after replacing the in-column solvent with the shipping solvent. Securely tighten the end caps and store the column at a location with stable temperature (a cool and dark space is recommended).
  • Page 8 3-3 If the Shodex product is disposed of 3-4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K.K. 3-5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below: a.

This manual is also suitable for:

Shodex odp2 hp-2dShodex odp2 hpg-2aF7622004F7622005F6714011