Operation Manual Shodex IC SI-50 4E Version 211004E Operation Manual Shodex™ IC SI-50 4E Sales Office: Technical Support: Showa Denko Europe GmbH info@shodex.de Shodex Business +49 (0)89 / 93 99 62 37 Konrad-Zuse-Platz 3 81829 Munich Quotes and Orders: Europe, Middle East, Africa, Russia Germany order@shodex.de...
Page 2
Operation Manual Shodex IC SI-50 4E Version 211004E Operation Manual Shodex IC SI-50 4E (Please read this manual carefully to achieve the best and consistent column performance for a long time) Important Handling Instructions Caution! • Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column and understand their proper handling methods to prevent potential health hazards or death from occurring.
Page 3
Operation Manual Shodex IC SI-50 4E Version 211004E Usable Conditions Maximum Temperature (°C) Flow Rate (mL/min) Product Pressure (MPa) Name Range Recommended Maximum Recommended Range (Per Column) IC SI-50 4E 3 - 12 20 - 60 IC SI-50G 3 - 12 20 - 60 Usable solvents are listed below.
Page 4
Operation Manual Shodex IC SI-50 4E Version 211004E Column Usage Procedure 7.1. HPLC System Preparation Wash entire LC system prior to the column installation, including all flow-lines and sample loop by switching the valve, and then replace the washing solution with the eluent to be used. If desired new eluent has low miscibility/solubility to the eluent of previous analysis, first use the eluent that is miscible/soluble to both eluents, and then replace it with the desired eluent.
Page 5
Operation Manual Shodex IC SI-50 4E Version 211004E 7.3. Solvent Exchange When replacing the solvent, start the system at less than 0.3 mL/min. Recommended solvent volume to introduce at each step is 3 to 5 times of the column volume.
Page 6
Operation Manual Shodex IC SI-50 4E Version 211004E Column Storage Remove the column from the system after replacing the in-column solvent with the shipping solvent. Refer to section 7.3 Solvent Exchange for how to replace the eluent. Securely lighten the end caps and store the column at a location with stable temperature (a cool and dark space is recommended).
Page 7
3-3 If the Shodex product is disposed of 3-4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K.K. 3-5 If the performance of the Shodex product is not conform to the specification of the attached...