Service - Maintenance
Reassembly
3
2
1 Storage box
2 Work plan
1. Place the work plan vertically on the frame.
2. Check that the control panel cable is clear: the control panel must be able to move at least 1 m
3. Pull the work plan towards the front of the detector then clip it on.
4. Refit the flange around the inlet pipe.
5.4 Front cover disassembly/reassembly
Prerequisite for disassembly
► Disassemble the work plan (see chapter "Work plan disassembly/reassembly").
Disassembly
1 Latch
1. Operate each latch to unlock the cover.
2. Pull the cover forwards.
Prerequisite for reassembly
► Power off the leak detector (see chapter "Powering Off").
14/34
4
1
3
Inlet pipe
from its support.
2
Cover
1
1
2