Notes On Reprocessing Of Resterilizable Products - Leica M844 User Manual

Hide thumbs Also See for M844:
Table of Contents

Advertisement

11.6.2 Ottoflex
• Use only precision-formed original Leica 12 V/50 W
halogen lamps.
• Never touch the glass bulb of halogen lamps with your
bare fingers.
Open the cover on the rear of the optics carrier by sliding the
X
lamp cover opener to the right.
1
2
3
1 Reflector
2 Bulb socket
3 Lever
4 Bulb slider
5 Lamp cover opener
Make sure that the quick-change lamp mount is on Main Light 1
X
so that the bulb slider (1) is all the way to the right.
Fold the reflector (3) on the pin upwards and to the right.
X
Pull on the tab to remove the bulb socket (4) complete with
X
bulb.
Insert the new bulb socket (4) and bulb.
X
Fold the reflector back down until it engages.
X
Whenever you replace a bulb, reset the bulb's hour meter to
0 (see chapter 9.6.1)
4
5
M844 / Ref. 10 713 294 / Version 01
11.7 Notes on reprocessing of
resterilizable products
11.7.1 General
Products
Reusable products supplied by Leica Microsystems (Schweiz) AG
such as rotary knobs, objective protective glasses and capping
pieces.
Limitation of reprocessing
For the medical devices used on patients suffering from Creutzfeldt
Jacob Disease (CJD) or suspected of having CJD or variant CJD, the
local statutory requirements have to be met. Normally resterilizable
products used on this group of patients are to be eliminated
without risk by incineration.
Occupational safety and health protection
Particular attention must be paid to the occupational safety and
health protection of the persons responsible for preparing
contaminated products. Current regulations of hospital hygiene and
prevention of infection must be observed in the preparation,
cleaning and disinfection of the products.
Limitation of reprocessing
Frequent reprocessing has little effects on these products. The end
of the product life cycle is usually determined by wear and tear and
damage through use.
11.7.2 Instructions
Workplace
Remove surface contamination with a disposable cloth/paper
X
cloth.
Storage and transport
• No special requirements.
• It is recommended to perform the reprocessing of a product
immediately following its use.
Preparation for cleaning
Remove the product from the M844 surgical microscope.
X
Cleaning: manually
• Equipment: running water, detergent, alcohols, microfiber cloth
Procedure
Rinse surface contamination off of the product (temp. < 40 °C).
X
Use some rinsing agent depending upon degree of
contamination.
Alcohol may also be used to clean the optics if heavy
X
contamination such as fingerprints, grease streaks etc. is
present.
Care and maintenance