Preparation before surgery

7
Preparation before surgery
7.1

Transportation

WARNING
Danger of injury due to:
• uncontrolled lateral movement of the arm system,
• tilting of the stand,
• trapping of feet in lightweight shoes beneath the casing
of the base.
For transportation, always move the M844 surgical
X
microscope into the transport position.
Never move the stand while the unit is extended.
X
Never roll the stand or OP equipment over the cables
X
lying on the floor.
Always push the M844 surgical microscope; never pull it.
X
CAUTION
Surgical microscope can move without warning.
Always lock the footbrake when you are not moving the
X
system.
CAUTION
Danger of injury from uncontrolled lateral movement of the
system and arm systems.
When transporting or moving the microscope (F20, F40)
X
on an inclined plane, always lock the swing arm, monitor
arm and control unit (see 14.12)
When storing the microscope (F20 only) on an inclined
X
plane use the provided wedge from the transport box
(see 14.12).
NOTE
Damage to the M844 surgical microscope during transportation.
Never move the stand in the extended condition.
X
Never roll the stand or OP equipment over the cables lying on
X
the floor.
NOTE
Damage to the M844 surgical microscope due to uncontrolled
tilting.
Hold the handle when releasing the brake.
X
The M844 shall be moved only with released brakes.
Do not perform any movements when the brakes are
X
locked.
26
7.1.1
WARNING
Beware of:
CAUTION
Surgical microscope can move without warning!
Whenever you transport your Leica M844 F20, first bring it into the
transport position.
4
Unplug and secure the power cable.
X
Pull the stop lever (1) and turn it into a horizontal position.
X
Move the swing arm up and down until the stop lever engages.
X
Open the articulation brakes (2).
X
Bring swing arm into transport position.
X
Tighten the articulation brakes (2).
X
Turn the control unit towards the XY-unit.
X
M844 / Ref. 10 713 294 / Version 01
Transporting the M844 F20
• Uncontrolled lateral movement of the swing arm!
• Tilting of the stand!
• Feet in lightweight shoes could become trapped
beneath the casing of the base.
• Collision between user and microscope system. E.g.
between head and camera control unit holder
(CT40)
• Abrupt braking of the surgical microscope at a
threshold that cannot be crossed.
Before transport, always move the Leica M844 F20
X
surgical microscopes into the transport position.
Never move the stand in the extended condition.
X
Always push the surgical microscope; never pull it.
X
Never roll over cables lying on the floor.
X
Always lock the footbrake when you are not moving the
X
system.
2
3
1