7.9

Fitting the Leica Slit Illuminator

7.9.1
General safety information when handling
the Leica Slit Illuminator
• When installing and using the slit illuminator, make sure not to
pinch any cables.
• When installing, make sure that the interlock of the slit
illuminator latches securely.
• Only qualified personnel is allowed to handle the slit
illuminator.
• When handling the slit illuminator take care not to crush any
fingers.
7.9.2
Mounting the extension plate
Lock the swing arm.
X
Fasten the extension plate (2) to the optics carrier with
X
four screws (3).
3
1
2
M844 / Ref. 10 713 294 / Version 01
7.9.3
Fastening the slit illuminator
WARNING
Risk of injury from parts falling down!
Complete all preparations and adjustments to the optics
X
carrier before the operation.
Never rebalance or re-equip the optical components and
X
accessories with the instrument over the operation area.
Before the operation, make sure that the optical
X
components and accessories are installed properly.
Before re-equipping during the operation, first swing
X
the microscope away from the operating field.
Before each operation, make sure that the optical
X
components and accessories are sufficiently secured and
cannot move.
Make sure that the interlock latches securely.
Loosen the clamp screw (1) and slide the slit lamp (5) into the
X
guide (6).
Tighten the clamping screw (1).
X
Power supply and control signals are connected to the slit
illuminator via the quick release fastener in the guide (6).
6
The slit illuminator may only be used with an objective
(10445937) with a working distance (WD) of 200 mm.
Insert the 3-pin plug of the dual cable into the external supply
X
socket (4) on the optics carrier.
Preparation before surgery
4
5
33