07-Start Key and Led:
This key starts the centrifugation operation and the led turns on. The led turns off if the
program is stopped manually or the lid is opened at the end of the program.
08-Value Decreasing Key :
This key is used to decrease the parameter values during programming.
09-Stop key :
This key is used to stop the operation manually.

6.1 LOADING

 Each tube insert must be at the same weight as the one diametrically opposite for balancing.
 In case the number of tubes which will be centrifuged is less than the capacity of the rotor, the
tubes must be placed in opposite inserts. If the number of tubes that will be centrifuged is an
odd number, use a water-filled tube at the same weight for balancing.
Imbalance of the rotor may cause major damage to the rotor and centrifuge
Never attempt to introduce liquids into the tube inserts.
Balance the rotor with glass tubes if you use glass tubes for centrifugation.
Balance the rotor with plastic tubes if you use plastic tubes for centrifugation.

SECTION 6

OPERATING PRINCIPLES

RIGHT LOADING
WRONG LOADING
Figure 3
10

Hide quick links:

Advertisement

Table of Contents
loading