Environmental Specifications; Electrical Requirements - Siemens Clinitek Advantus Operator's Manual

Hide thumbs Also See for Clinitek Advantus:
Table of Contents

Advertisement

Environmental Specifications

  s
Specification
  n
   i   o
 Ambient Operating
   t
  a
Temperature
   i   c
   f
   i
  c
  e
  p
   S
Optimum Operating
Temperature
Relative Humidity
Optimum Relative
Humidity
Indoor Use Only
 Altitude
Ventilation
IEC 1010-1 Installation Category II
IEC 1010-1 Equipment Classification Class I
IEC 1010-1 Pollution Degree 2

Electrical Requirements

Requirement
Electrical Rating
Power Requirements
Maximum Power Input
Fuse Rating
Line Leakage Current
124
Value
18°C to 30°C (64°F to 86°F).
 At temperatures under 22°C (72°F), urobilinogen
and leukocyte results may be decreased, and at
temperatures above 26°C (79°F), increased.
22°C to 26°C (72°F to 79°F).
20% to 80%, non-condensing, actively controlled
35% to 55%
up to 2000 meters
1709 BTU
Value
100 to 240 VAC
10%
±
50 to 60 Hz
72 VA
2 A, 250 V, 2 AG, SB(T)
<0.5 mA in normal condition
<3.5 mA in single fault condition
Clinitek Advantus Operator's Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents