5. Remove the fixed platform by pulling the entire assembly towards you.
  g
   i   n
   t
  o
  o
   h
  s
   l   e
   b
  u
  r   o
   T
Figure 20 Remove the Fixed Platform
6. Remove the holddown plate from the fixed platform. Refer to Performing 
the Daily Cleaning  ‚ page 37.
7. Remove the jammed strip.
8. Reinstall the holddown plate. Refer to Performing the Daily Cleaning  ‚
page 37.
9. Reinstall the fixed platform. Refer to Performing the Daily Cleaning  ‚
page 37.
10. Reinstall the push bar. Refer to Performing the Daily Cleaning ‚ page 37.
11. Turn analyzer power on.
12. Rerun the specimen(s) without results.
56
Clinitek Advantus Operator's Guide