Iq sensor net, valve module for compressed air-operated sensor cleaning head
Table of Contents

Advertisement

MIQ/CHV PLUS
 Small, clamping range 4.5 to 10 mm. This cable gland is suitable for all
 Large, clamping range 7 to 13 mm. This cable gland is required for cable
General
Observe the following points when attaching connecting wires to the terminal
installation
strip
instructions
 Shorten all wires to be used to the length required for the installation
 Always fit all the ends of the wires with wire end sleeves before connecting
 Any wires that are not used and project into the enclosure must be cut off as
 Screw a small cable gland with sealing ring into each remaining free opening
ba75539d03
03/2018
All manuals and user guides at all-guides.com
IQ S
N
sensor cables.
ENSOR
ET
sealing ring 20 x 15 x 1 mm
cable gland M16
blind plug
sheaths with an outside diameter of more than 10 mm and is screwed into the
enclosure via an extension piece.
sealing ring 20 x 15 x 1 mm
extension piece M16/M20
sealing ring 24 x 19 x 2 mm
cable gland M20
If necessary, you can order more large cable glands in a set of 4
pieces (Model EW/1, Order No. 480 051).
them to the terminal strip
closely as possible to the cable gland.
and close it with a blind plug.
Installation