Figure 5
Electrical connections PMA 45K1
14.3
DETERMINATION OF THE TWO MEASURING RANGES TO BE SELECTED ON
THE INSTRUMENT
In order to minimize operating errors, the PMA45K1 provides the possibility to determine two measur-
ing ranges that may then be selected via a switch on the instrument.
The selection of measuring range 1 and 2 is made via two jumpers on the below illustrated plate. On
the photo, you can see that for range 1 the 1 % measuring range has been selected and for range 2
the 2.5 % measuring range. The selected measuring ranges refer only to the mA-outlet. The digital
indication is always in the 100.00 vol% O
Via the external measuring range selection, all measuring ranges are available.
www.mc-techgroup.com
measuring range.
2
PMA45K1 | 1.00.01
15