Crvhdr? Curve Header Query; Crvpt Curve Data Point Command; Crvpt? Curve Data Point Query - Omega CYD211 User Manual

Single input temperature monitor
Table of Contents

Advertisement

Curve Header Command (Continued)
Example:
CRVHDR 21,DT-470,00011134,2,325.0,1[term] –
Configures User Curve 21 with a name of DT-470, serial
number of 00011134, data format of volts versus Kelvin,
upper temperature limit of 325 K, and negative coefficient.
CRVHDR?
Curve Header Query
Input:
CRVHDR? <curve>[term]
Format: nn
<curve>
Returned: <name>,<SN>,<format>,<limit value>,<coefficient>[term]
Format: aaaaaaaaaaaaaaa,aaaaaaaaaa,n,+nnn.nnn,n
(Refer to command for description)
Remarks:
Returns a standard or user curve header.
CRVPT
Curve Data Point Command
Input:
CRVPT <curve>, <index>, <units value>,
<temp value>[term]
Format:
nn,nnn,±nnnnnnn,+nnnnnnn
<curve> Specifies which curve to configure. Valid entry:
<index> Specifies curve points index. Valid entries: 1 – 200.
<units value> Specifies sensor units for point to 6 digits.
<temp value> Specifies the corresponding temperature in
Remarks:
Configures a user curve data point. To finalize curve entry,
send the *RST command or cycle the instrument power
after all the curve points have been entered.
Example:
CRVPT 21,2,0.10191,470.000[term] – Sets User Curve 21
second data point to 0.10191 sensor units and 470.000 K.
CRVPT?
Curve Data Point Query
Input:
CRVPT? <curve>, <index>[term]
Format: nn,nnn
<curve>
<index>
Returned: <units value>, <temp value>[term]
Format: ±nnnnnnn,+nnnnnnn (Refer to command for description)
Remarks:
Returns a standard or user curve data point.
4-18
Omega Model CYD211 User's Manual
Valid entries: 1 – 21.
21.
Kelvin for this point to 6 digits.
Specifies which curve to query: 1 – 21.
Specifies the points index in the curve: 1 – 200.
Remote Operation

Advertisement

Table of Contents
loading

Table of Contents