pH
% S L OP E
View
clear
4
7
10
BUFFER
BUFFER
2
4
7
10
12
select
select
°C
pH
S T AB L E
View pH electrode slope
- View the pH electrode slope in
% after 2-point standardization
is successfully done.
Clear pH standardization
- To clear pH standardized
buffers.
pH buffer group
- Select from 3 different buffer
groups, each containing 5-6
buffers, for auto buffer
recognition.
Temperature Unit
- Select unit of measure for
Temperature either in °C or °F.
pH resolution
-
Select various pH resolution
either 88.888, 88.88 or 88.8.
(0.001/ 0.01/ 0.1)
Stability Indicator
-Set the stability indicator to be
displayed on the screen. Select
On or OFF.
Overview of Setup Menus in AB15+
10
P rint
Print
Print
Print
clear
Baud Rate
- To set the baud rate
(bits of second) of the
communication protocol
interface. Select from 4800,
9600, 19200 or 38200.
Parity Bit
- Set the parity bit of the
communication protocol
interface. Select 0(none),
1(odd) or 2(even).
Stop Bit
- Set the stop bit of the
communication protocol
interface. Select 1 or 2.
Print Data
- To print current displayed
data or data stored in the
meter's memory to a
computer or printer via
Rs232.
Clear Stored Data
- To clear all stored data sets
for new data to be stored.