Example: Making High Power Measurements With Option H85

Final Setup

21.Verify that all of the power and attenuator settings are correct.
22.In the Response drop-down menu select Measure > S21.
Figure 31
Response Menu
23.To Perform a response calibration following steps a, b, c and d:
a. Connect the test port cables of the analyzer to form a thru configuration.
b. In the Response drop-down menu select Cal Wizard (see
select UNGUIDED Calibration (Response, 1-port 2-port): Use Mechanical Standards >
Next.
Figure 32
Calibration Wizard
34
Figure
31). In the dialog box,
N5245-90002 User's and Service Guide