Keysight N5244/45AS Option H85 User's And Service Manual page 31

Table of Contents

Advertisement

To find the User Preset:
2. In the Utility drop-down menu select User Preset. Check User Preset Enable >
Save current state as User Preset and press OK.
Figure 23
User Preset
3. In the Stimulus drop-down menu select Power > Power and Attenuators. In the dialog box
type [–20 dBm] into Port Power for Port 1. Verify that Port Powers Coupled is checked to
ensure that Ports 1 and Port 2 power levels are the same. Press OK. Uncoupled ports should
be used when adjusting the S12 power level. 
Figure 24
Power and Attenuators
N5245-90002 User's and Service Guide
Example: Making High Power Measurements With Option H85