Keysight N5244/45AS Option H85 User's And Service Manual page 22

Table of Contents

Advertisement

N5244/45AS Option H85
Operation
In this configuration the DUT supplies the power. The measure receiver attenuator for B IN may
have sufficient range to compensate for the users desired power level. Port 2 has a high power
isolator located between the SOURCE OUT and CPLR THRU. This isolator is required to protect
the components inside the PNA-X. The isolator can be replaced by a high power attenuator or
circulator. Reference Channel R1 can be set to internal. Refer to
page
19.
Figure 18
2-Port Forward DUT Configuration
High Power Forward
DUT Configuration
18
DUT
Figure 18
and
Figure 19 on
RCVR B IN
CPLR ARM
Fix Attenuator
SOURCE OUT
CPLR THRU
Out
In
Isolator
N5245-90002 User's and Service Guide