4

Maintenance and Troubleshooting

4.1

Maintenance

Under normal operating conditions, no regular maintenance is required except for occasional cleaning of
the front panel and replacement of the battery (option NRT-B3).
4.1.1

Cleaning the Outside

For cleaning the outside of the instrument it is recommended to use a soft, lint-free cloth and a non-a-
lcoholic solvent, eg a commercial detergent.
Note:
In no case use solvents such as nitro thinners, acetone or similar, since these may damage
the front-panel labels or plastic parts.
4.1.2

Replacing the Battery (Option NRT-B3)

Removal of battery
Fig. 4-1
Insertion and removal of battery
The NiMH battery used in the NRT is a high-performance battery with a lifetime of typically more than
200 charge/discharge cycles.
The battery should be replaced if the operating time with a Sensor NRT-Z44 is less than 4 hours. Spare
batteries are available from Rohde & Schwarz under the designation NRT-Z1 (Order No. 1081.1209.02).
The battery comes complete with the plastic cap used as a holder.
For batteries from the manufacturer FEDCO (type designation ENERGY+ DR30AA), the plastic cap of
the old battery has to be used (carefully lift the locking flaps when removing it).
1080.9506.02
Press
Withdraw
Press
4.1
Maintenance
Insertion of battery
Press in center
E-4