Nitrogen N*10DR Nitrogen Generator UM – N*10DR – Rev 4 History Document Change History Rev No. Change Initials Date First Release G.L. 16.10.06 Notes added for - Indoor use only & Altitude T.G. 10.09.07 Update USA address 25/08/2008 Updated solenoid valve part # 26/01/2009...
These instructions must be read thoroughly and understood before installation and operation of your Peak N*10DR generator. Use of the generator in a manner not specified by Peak Scientific MAY impair the SAFETY provided by the equipment. When handling, operating or carrying out any maintenance, personnel must employ safe engineering practices and observe all relevant local health and safety requirements and regulations.
Nitrogen N*10DR Nitrogen Generator UM – N*10DR – Rev 4 Technical Specification Generator Environment Minimum operating ambient temperature 5° C (41° F) Maximum operating ambient temperature 25° C (77° F) Maximum relative humidity Maximum Altitude 2000 Meters Nitrogen Outlet Maximum flow 30 l/min (1.1 CFM) Maximum Pressure 6.89 bar (100 psi)
Thank you for selecting Peak Scientific to meet your gas generation needs, and should you require any further assistance or support please do not hesitate to contact...
Installation Unpacking and Installation Although Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any sign of transit damage. Check ‘SHOCKWATCH’ label for signs of rough handling prior to unpacking -...
Nitrogen N*10DR Nitrogen Generator UM – N*10DR – Rev 4 Installation Removal of Transit Bracket Remove Rear Panel as follows:- At the rear of the unit, turn the key in the lock anti-clockwise to unlock the rear panel. Hold the rear panel by the two handles and slowly swivel the bottom of it out and lower it until the top comes free from the chassis.
Nitrogen N*10DR Nitrogen Generator UM – N*10DR – Rev 4 Installation Useful Information The diameter of the tubing which will be connected to the gas outlets is important and is determined by the length of tubing required. Failure to follow these recommendations could lead to accelerated compressor wear.
Nitrogen N*10DR Nitrogen Generator UM – N*10DR – Rev 4 Generator Environment The Generator is designed for indoor use only. The unit should be installed adjacent to the Mass Spectrometer it is supplying. If this is not convenient then the unit can be sited elsewhere, however, consideration should be made of the lengths of pipe runs as pressure drops can result from extended runs of pipe.
Nitrogen N*10DR Nitrogen Generator UM – N*10DR – Rev 4 Connections to Rear of Unit Connections Serial Plate Power Cord (230V ONLY) 04-1038 (EU) 04-1029 (US) Water outlet 04-1030 (UK) 1/4” BSPP female Rear panel lock 9-way D-type connector for Nitrogen outlet PC Comms.
The display will read - Peak Scientific Mode: APCI *Note: The outlet should be vented to atmosphere for at least 30 seconds to purge the system of impurities.
21 days time: 30 days System Hours To exit. MENU ENTER To display the software revision. System Hours MENU Software Rev. V1.55 To exit. MENU ENTER Mode: To display the current operating mode. Mode Peak Scientific ENTER MENU Mode: ESI Diagnostic...
•Rotate knurled locking nut at base of adjustment knob to lock. Unusual Operation If at any time the generator begins to emit excessive noise or vibration, then it should be switched off and you should contact your local representative or Peak Scientific as soon as possible. System Drain Please ensure that the drain port at the rear of the generator is led to a suitable connection or container make sure the container is not sealed tight.
Nitrogen N*10DR Nitrogen Generator UM – N*10DR – Rev 4 Routine Maintenance Maintenance WARNING Servicing and/or repair of the Generator should only be undertaken by a TECHNICALLY COMPETENT PERSON, with the generator in its safely isolated condition. SAFELY ISOLATED CONDITION Definition: The unit is in a Safely Isolated Condition when it is disconnected from its application, fully de-pressurised and isolated from the Electrical Supply.