Additional Notes On Limit Testing - HP 8712C User Manual

Rf network analyzers
Table of Contents

Advertisement

Stimulus and Amplitude
In frequency sweep mode, the stimulus values are interpreted as frequencies;
Values
in power sweep mode, the stimulus values are interpreted as output power
levels.
The values entered for stimulus and amplitude are unitless. If the
C A U T I O N
measurement format is changed after limit lines are set, the amplitude values
do not automatically change. Therefore, you should select the format you
want to use before entering your limits.
Calculating the
Frequency of a Data
Point
The frequency of each data point is calculated using the following formula:
Freq @t-number) =
where pt-number ranges from 1 (for the leftmost display point) to
Limit testing is only performed on actual data points, not interpolated values
limit Testing and
Measurement Points
between them. The limit lines or points that you enter are converted
to values at each measurement point. Most of the time, this has little or
no effect on the validity of the limit testing. However, note the following
examples of problems that could arise when using limit lines:
(pt-number -
x (stop-freq - start-freq)
1)
number-of -points -
Using Instrument Functions
Using limit Testing
+ s t a r t - f r e q
1
4-47

Advertisement

Table of Contents
loading

This manual is also suitable for:

8714c

Table of Contents