Model 4262A
Section III
Paragraphs 3-52 and 3-53
(7)
COMPARATOR
ENABLE Switch: This
switch enables the Option 004 to compare
measured data with HIGH and LOW limits
under a fixed range condition (LCR or D/Q
RANGE switch set to MANUAL). If LCR
RANGE switch or D/Q switch is set to
AUTO, depressing COMPARATOR
EN¬
ABLE switch changes LCR or D/Q RANGE
switch setting to MANUAL.
If AUTO key of LCR or D/Q RANGE
switch is depressed while COMPARATOR
ENABLE switch is ON, one measurement
cycle is done in AUTO ranging and the
range is fixed to that selected in this meas¬
urement cycle.
3-52.
LIMIT Setting
Warning:
If HIGH
LIMIT setting is lower than LOW LIMIT setting,
HIGH
and
LOW
lamps
of
decision
output
repeatedly turn ON and OFF to warn operator to
change LIMIT setting.
3-53. DATA OUTPUT Connector Decision Output:
Decision outputs in TTL open collector signal and
in relay contact are supplied through COMPARA¬
TOR OUTPUT connector on the rear panel. Signal
pin assignment is given in Figure 3-44.
WARNING I
DO NOT APPLY AC LINE VOLT¬
AGE TO RELAY OUTPUT CON¬
NECTOR PIN TO SWITCH LINE
CURRENT.
For such relay appli¬
cations, remotely control an ex¬
ternal relay with relay output.
Relay Contact Ratings
AC
DC
Contact Resistance
lOOmfi
lOOmft
Maximum Permissible
Power
30VA
20W
Maximum Permissible
Voltage
110V
30V
Maximum Permissible
Current
0.3A
1A
Actuation Life
> 10 million
> 1 million
Decision Output Data Format
Decisions
Relay output pins
TTL output pins
DQ LCR
13
i?
DQ LCR
14
18
DQ LCR
39
43
DQ LCR
15
19
DQ LCR
16
20
DQ LCR
41
45
HI
s
o
0
H
L
L
IN
0
0
s
L
L
H
LO
0
s
0
L
H
L
S: Short
O: Open
Referenced to common (pin 38 or 42).
TTL Output sink current: 30mA max.
3-31