INSTALLATION
2

2.9 Mounting

2.9.1 Torques and pressures sandwich versions

WARNING!
Use stainless steel A2 / 6.9 class bolts.
Make sure the connecting flanges are of type raised face (RF).
Figure 2-18: Tighten the bolts in fixed order, see picture.
Max. torque:
• Step 1: approx. 50% of max. torque
• Step 2: approx. 80% of max. torque
• Step 3: 100% of max. torque given in tables
EN 1092-1
Nominal size
DN [mm]
ASME B 16.5
Nominal size
[inch]
14
Pressure rating
25...80