5.3 Replacing a Side Bolster (Pro)

Refers to spare parts kits: SP1633209, SP1633210
Wire clippers
1.
Switch off the control unit and disconnect from
mains supply.
2. Open the zipper of the bed cover and take off
the top cover (see Fig. 5-1).
3. Unplug all twenty-two snap fasteners connecting
the air cells to the defective side bolster.
4. Using the wire clippers, cut cable tie A securing
the upper L-shaped blue tube, connecting the
head cell to the side bolster and unplug it. Repeat
this procedure for the remaining two head cells
(there are three in total).
5. Unplug the four snap fasteners that secure the
side bolster to the mattress' bottom cover.
6. Discard the defective side bolster.
7. Install the new side bolster in place by following,
in reverse order, the process described on points
3. to 5. Secure the tubes with the provided cable
ties.
Cut the excess cable tie
8. Put the top cover back in place and close the
zipper (see Fig. 5-5).
60129675-B
length.
A
Fig. 5-8
Invacare
®