4. Facing towards the mattress, on the left-hand
side, two L-shaped blue tubes connect the air
cell to the bolster on that side. Using the wire
clippers, cut both cable ties A securing them.
5. Carefully cut the end of the L-shaped tube
connected to the side bolster using the wire
cutters and unplug it.
6. Remove the defective air cell by sliding it out of
the mattress and discard it (Fig. 5-4).
7. Install the new air cell in place.
8. Slightly brush off any existing dust from the
side bolster's holder and plug the the new cell's
L-shaped tubes into their respective holders.
9. Secure the tubes with the provided cable ties.
Cut the excess cable tie length.
10. Plug the snap fasteners on both sides to fix the
air cell in its place.
11. Put the top cover back in place and close the
zipper (Fig. 5-5).