7. Unplug the yellow earth wire C from its socket.
8. Remove the nonfunctioning compressor and
install the new one.
9. Install the six provided screws and tighten them.
10. Plug the compressor cable and the earth wire.
11. Install the air tube and secure it with the provided
cable tie. Cut the excess cable tie length.
12. Reinstall the housing. Ensure no cable or tube is
kinked or clamped.

4.2.5 Replacing the Motor

Refers to spare parts kit SP1631446
Risk of damaging the PCB by electro-
static discharge
Handle PCB's only in Electrostatic Protected
Areas
T10, T20 Torx keys
Wire clippers
1.
Switch off the control unit and disconnect from
mains supply.
2. Detach the top housing, the speaker cable and
both FPC's (see 4.2.2).
3. Remove the three air tubes A, B, and C.
4. Unplug the power switch cable and the motor
cable from the PCB (see 4.2.3).
5. Loosen and remove the two screws D with the
appropriate Torx key.
6. Remove the motor.
7. Install the new motor and screws. Tighten the
screws.
8. Plug the air tubes A, B, and C, and the cables
into the PCB.
9. Reinstall the housing. Ensure no cable or tube is
kinked or clamped.