4.1.6 Replacing the Air Tube Connector

Refers to spare parts kit SP1631535
Risk of damaging the PCB by electro-
static discharge
Handle PCB's only in Electrostatic Protected
Areas
T10, T20 Torx keys
1.
Switch off the control unit and disconnect from
mains supply.
2. Detach the top housing, the speaker cable and
both FPC's (see 4.1.2).
3. Hold the connector assembly firmly with one
hand. Remove all the air tubes A, B and C from
the connector.
4. Loosen and remove the screws D with the appro-
priate Torx key.
5. Discard both the nonfunctioning connector and
the bracket.
6. Install the new ones and tighten them with the
two provided screws.
7. Reattach the air tubes to the connector. Ensure
to install the tubes in the correct sequence.
8. Place the connector assembly on the bottom
housing. Ensure that tube C is the one at the
bottom.
9. Reinstall the top housing. Ensure no cable or
tube is kinked or clamped.