Cleaning Air Box
3
3.1
The following parts are included in the scope of delivery of the Cleaning Air Box:
Cleaning Air Box
Accessory set, including
Operating manual.
3.2
WARNING
The Cleaning Air Box is not suitable for operation in a potentially explo-
sive environment.
Ambient
If the Cleaning Air Box is installed outside it should be protected by a sun shield
conditions
against the effects of the weather (snow, ice and direct solar radiation). The
measurement location must meet the environmental conditions specified in
chapter 8 T
3.3
Electrical equipment (e.g. motors, contactors, cables, lines, relays) must meet
the following requirements:
Conformity with national regulations (e.g. NEC, VDE and IEC)
Suitability for the electrical conditions at the installation site
Suitability for the environmental conditions at the installation site
Suitable fusing of the electrical circuit
Suitable disconnecting device (e. g. switch or circuit breaker) for the mains
ba75731e04
12/2013
All manuals and user guides at all-guides.com
Installation
Scope of delivery
– Intake air filter
– 4 x screws M5x16
– Fitting for compressed air hose
– Hose clip
Requirements of the measurement location
.
ECHNICAL DATA
Safety requirements on the electrical installation
– Maximum operating voltage
– Maximum operating current
– Temperature resistance (minimum and maximum temperature)
– Stability against UV light when used outdoors
– Protection against water and dust (Nema or IP protection class).
– Overcurrent fuses (in accordance with the technical data of the device
input or output)
– Overvoltage class II surge limiters
supply of permanently mounted equipment with separate mains connection
Installation
3 - 1