Saros 4000 Operator's Manual page 36

Oxygen system
Table of Contents

Advertisement

Recommended separation distances between portable and mobile RF communications equipment and the SAROS
The SAROS is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the
SAROS can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equip-
ment (transmitters) and the SAROS as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
Classification
Type of protection against electric shock:
Class II
Protection from electric shock is achieved by
double
insulation.
Degree of protection against electric shock:
Type BF
Equipment providing a particular degree of
protection against electric shock regarding
1) allowable leakage current;
2) reliability of protective earth connection (if present).
Not intended for direct cardiac application.
Independent testing for Medical Electrical Equipment Standard:
Tested to be in compliance with
• IEC 60601-1 edition 3.1: Medical Electrical equipment -
Part 1 General Requirements for safety.
• CAN/CSA C22.2 No. 60601-1-8 edition 2.1, Medical Elec-
trical Equipment – Part 1: General Requirements for Safety
36 | MN255-1 C
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
m
80 MHz to 800 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
Protection against potential electromagnetic or other interference
between the equipment and other devices.
• Tested to be in compliance with EN 60601-1-2/IEC
60601-1-2: 4th Edition.
• Tested to be in compliance with RTCA/DO160
Section 21 Category M.
• CISPR 11 / EN 55011 Class B Group 1, "Industrial,
Scientific, and Medical (ISM) Equipment"
• FCC Part 15, Subpart B – Class B Unintentional Radiators
IP33 – Protection against ingress of tools, thick wires, etc. >
2.5mm. Water falling as a spray at any angle up to 60° from the
vertical shall have no harmful effect.
Method of cleaning and infection control allowed: Please refer to
cleaning section of this SAROS manual.
Degree of safety of application in the presence of flammable anes-
thetic gases: Equipment not suited for such application.
Mode of operation: Continuous duty.
Operator Manual
800 MHz to 2,5 GHz
d
1
2 .
P
0.23
0.73
2.3
7.3
23