Pulse Settings
autoSAT Technology
Trigger Sensitivity
Trigger Criteria
Minimum time between breaths
Response to Missing Breaths
NOTE: At breath rate above Max Breath Rate, bolus size will be reduced proportionally.
Input Voltage
Input Current
Output Voltage
Output Power
Output Voltage
Capacity
Battery Life
Battery Recharge Time
16, 32, 48, 64, 80, 96 ml
Servo-control to maintain consistent FiO2
Adjustable settings of 1 (most sensitive), 2 and 3 (least sensitive)
Cannula pressure has dropped below the trigger point (typically between 0.15 – 0.45 cm of H2O of
negative pressure with a MAX trigger point of .50 cm H2O)
1.25 seconds (max 3 consecutive breaths)
Switch to Continuous Mode if no inspiration has been detected for 60 seconds with an audible
alarm.
Table 1-2. Pulse Mode Specifications
Bolus Size ml (± 15%)
16
32
48
64
80
96
Table 1-3. Bolus Volume and Breath Rates
AC Power Supply
100- 240 VAC, 50 – 60 Hz
115V at 400Hz
2.5- 1.3A
24 VDC
200 W
Table 1-4. Power Accessory Specifications
14.6 VDC
87.6 W-Hr
80% Nominal Capacity after 200 Charge/Discharge cycles
1.5 hours typical (3 hours max) to achieve 80% capacity from a fully discharged battery
while operating at 3 lpm
Table 1-5. Battery Specifications
Max Breath Rate
40
40
40
40
37
31
24 VDC Cable
20 – 28 VDC; 24VDC (Nominal)
10 A
-
-
MN255-1 C | 11