Emc Testing - Saros 4000 Operator's Manual

Oxygen system
Table of Contents

Advertisement

Guidance and Manufacturer's Declaration —Electromagnetic Emissions
The SAROS is intended for use in the electromagnetic environment specified below. The customer or the user of the SAROS should assure that it is
used in such an environment.
Emissions Test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Guidance and Manufacturer's Declaration—Electromagnetic Immunity
The SAROS is intended for use in the electromagnetic environment specified below. The customer or the user of the SAROS should assure that it is
used in such an environment.
Immunity test
IEC 60601 test level
Electromagnetic environ-
±8 kV contact
ment – guidance
±15 kV air discharge
IEC 61000-4-2
±2 kV for power supply lines
Electrical fast transient/
burst IEC 61000-4-4
±1 kV for input/output lines
±1 kV line(s) to line(s)
±2kV common mode on
Surge
AC lines
IEC 61000-4-5
±1kV differential on AC lines
±2kV common mode on
outdoor I/O lines
0% U
90°, 135°, 180°, 225°, 270°,
315°)
Voltage dips, short
interruptions and voltage
0% U
variations on power
supply input lines IEC
70% U
61000-4-11
25/30 cycles (0
0% U
Power frequency (50/60
Hz) magnetic field IEC
3 A / m
61000-4-8
NOTE U
is the A.C. mains voltage prior to application of the test level.
T
NOTE In some cases, the SAROS may reset after a significant Electrostatic Discharge, and may require a Full Power Cycle to resume normal opera-
tion. See Troubleshooting section for more details.
34 | MN255-1 C

10.0 EMC TESTING

Compliance
Electromagnetic Environment - Guidance
The SAROS uses RF energy only for its internal function. Therefore, its RF emis-
Group 1
sions are very low and are not likely to cause any interference in nearby electronic
equipment.
Class B
The SAROS is suitable for use in all establishments, including domestic establish-
Class A
ments and those directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Complies
Compliance level
±8 kV contact
±15 kV air discharge
±2 kV for power supply lines
N/A
±1 kV line(s) to line(s)
for 0.5 cycle (0°, 45°,
0% U
T
90°, 135°, 180°, 225°, 270°,
315°)
for 1 cycle (0
)
0% U
0
T
(30% dip in U
) for
70% U
T
T
0
)
25/30 cycles (0
for 250/300 cycles (0
)
0% U
0
T
3 A / m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of a typical
professional healthcare facility and home health-
care environments.
Mains power quality should be that of a typical
professional healthcare facility and home health-
care environments.
for 0.5 cycle (0°, 45°,
T
Mains power quality should be that of a typical
professional healthcare facility and home health-
care environments. If the user of the SAROS
for 1 cycle (0
)
0
T
requires continued operation during power mains
interruptions, it is recommended that the SAROS
(30% dip in U
) for
is powered from an uninterruptible power supply
T
T
0
)
or a battery.
for 250/300 cycles (0
)
0
T
Power frequency magnetic fields should be at levels
characteristic of a typical location in a professional
healthcare facility and home healthcare environ-
ments.