Alarm Indications And Codes - Saros 4000 Operator's Manual

Oxygen system
Table of Contents

Advertisement

Operator Manual
Priority #
Alarm Signal
1
PN
2
HU
3
HB
4
HU
5
HU
6
HU
7
SM
8
SE
9
SE
10
SM
11
SE
12
SE
13
SE
14
SE
15
SE
16
PN
17
SE
18
TN
19
TN
20
BL
21
OL
22
XF
23
XB
24
XK

6.2 ALARM INDICATIONS AND CODES

Alarm Type
Alarm Code
Power None
4000
Hot Unit
9510
Hot Battery
9300
Hot Unit
9200
Hot Unit
9100
Hot Unit
9110
Shutdown
A500
Mechanical
Shutdown
AC00
Electrical
Shutdown
8000
Electrical
Shutdown
9194
Mechanical
Shutdown
9195
Electrical
Shutdown
A300
Electrical
Shutdown
A000
Electrical
Shutdown
A010
Electrical
Shutdown
9000
Electrical
Power None
4010
Shutdown
FD00
Electrical
Trigger None
B000
Trigger None
B010
Battery Low
9540
Oxygen Low
0800
Malfunction
2000
Flow
Malfunction
9500
Battery
Malfunction
8110
Keypad
Alarm Description
Loss of Power
Battery Too Hot
Motor Driver Board Too Hot
Power Board Too Hot
Compressor Too Hot
Compressor Temperature Sensor
Failure
Unable to Achieve or Maintain
Product Tank Pressure
External Voltage Too High
Invalid Reset
Compressor Stall
Compressor Driver Fault
Ultrasonic Sensor Failure
IPC Failure
Board Communication Failure
Watchdog Time out
Missing Power POST
Motor Controller Fault
No Breath Detected for 15-59
seconds
No Breath Detected for 60
seconds
Low Battery ≤ 6 minutes
Oxygen Concentration Low
(< 85% nominal)
Flow Rate Error
Battery Communication Error
Key Stuck On
Buzzer
1 beep only
1 beep every 20 seconds for 120 seconds;
LCD Battery Icon blinking
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds for 120 seconds
1 beep only
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds for 120 seconds;
Power LED blinking
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds for 120 seconds
1 beep only
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds for 120 seconds
1 beep only; Automatically switches to
continuous mode
1 beep every 20 seconds for 120 seconds;
LCD Battery Icon blinking
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds for 120 seconds
1 beep every 20 seconds until condition
clears
1 beep every 20 seconds until condition
clears
MN255-1 C | 31