Saros 4000 Operator's Manual page 24

Oxygen system
Table of Contents

Advertisement

PULSE FLOW MODE FEATURES:
The SAROS Pulse Flow Mode delivers a high flow oxygen bolus at the beginning of each inspiration. Your SAROS Pulse Flow Mode has a
feature, called autoSAT Technology and a Pulse Flow Mode Sensitivity Setting to easily trigger the oxygen bolus delivery.
These features are provided to ensure that patients can easily trigger a delivery of an oxygen bolus and that the oxygen bolus maintains
consistency over higher breath rates.
autoSAT Technology provides a consistent bolus volume of oxygen as a patients' breath rate increases or decreases.
NOTE: Bolus volume decreases as breath rate exceeds published range.
The Pulse Flow Sensitivity (P SENS) can be adjusted in the Utility Menu from 1–3. A setting of 1 is the most sensitive and a setting of 3 is
the least.
WARNING: PULSE FLOW MODE SETTINGS MUST BE DETERMINED FOR EACH PATIENT INDIVIDUALLY. SETTINGS USED IN THE
CONTINUOUS FLOW MODE MAY NOT APPLY, OR BE APPROPRIATE, IN THE PULSE FLOW MODE AND VICE VERSA.
WARNING: AS WITH ALL OXYGEN-CONSERVING TYPE DEVICES, THE SAROS MAY NOT BE ABLE TO DETECT ALL INSPIRATORY
EFFORTS IN THE PULSE FLOW MODE.
Step 5: ADJUST THE FLOW SETTING TO THE PRESCRIBED LEVEL
Using the Flow Setting Buttons, adjust the flow setting to the prescribed setting.
Step 6: SELECT USER DEFINABLE SETTINGS USING THE UTILITY MENU
The Utility Menu has the features that can be adjusted by the user or information for the user or Service Technician. Below are the features of
the Utility Button and the Utility Menu:
• The Utility Menu mode can be entered whenever the unit has power (Standby or ON) or 9V Battery.
• Each time the Utility Button is pressed, the Utility menu will advance to the next option.
• The options within the Utility Menu can be changed by using the Increase and Decrease buttons.
• Pressing the Flow Mode button in the Utility Menu goes back to the previous Utility menu item.
• The Utility Menu will revert back to the previous display mode after 5 seconds.
The following sequence of screens shall be displayed on the LCD when the Utility Menu is displayed.
24 | MN255-1 C
Pulse Setting
Bolus Size (+/- 15%)
16
32
48
64
80
96
Table 4-3. Pulse Settings
Maximum Breath Per Minute
(bpm)
40
40
40
40
37
31
Operator Manual