Run Device And Fully Drain Saros Battery; Replacing The 9V Battery; Alerts, Alarms, And Troubleshooting Table; Figure 5-5. Replacing 9V Battery - Saros 4000 Operator's Manual

Oxygen system
Table of Contents

Advertisement

Operator Manual
Operate the device once every 3 months by running the device on battery power until drained. This is required if the device is in storage
and unused. If the device is not run regularly per this recommendation there is a risk it will not perform to specification when removed from
storage.
1) Turn on the SAROS and run at any continuous flow setting until the battery is completely drained and the SAROS powers off.
2) Once the SAROS powers off, immediately connect to AC external power to begin re-charging the battery. You may either continue to run the
SAROS or leave it powered off during re-charge.
3) Once the battery is fully charged, disconnect the battery, power off the unit if still running, and return the SAROS to proper storage conditions.
5.6.1 Disconnect power supplies and remove external battery if connected.
5.6.2 Using a Phillips #1 screwdriver, loosen the four captive screws shown in figure 5-2.
5.6.3 Remove cover.
5.6.4 Remove 9V battery from where shown in figure 5-5.
5.6.5 Disconnect and replace battery as shown in figure 5-5.
5.6.6 Using a Phillips #1 screwdriver, replace the Air Inlet Cover.
5.6.7 Initial and date the Service and Maintenance Record.
The tables show the possible audible and visual alerts and alarms, their conditions and suggest appropriate troubleshooting responses. The
SAROS runs functional verification on alarm system as part of its operation. If you are unsure about any alerts or alarm conditions, or
responses please contact an authorized service technician or go to www.caireinc.com.
Alarm levels
Type
Level
Alarm
Status
WARNING: AVAILABILITY OF A BACK-UP SOURCE OF OXYGEN IS RECOMMENDED IN CASE OF POWER OUTAGE OR A DEVICE
FAILURE. CONSULT YOUR PROVIDER FOR BACK-UP OXYGEN SYSTEM.
WARNING: DO NOT Ignore Alarms. System will attempt to produce oxygen under fault condition but may not be at purity
or flow.

5.5 RUN DEVICE AND FULLY DRAIN SAROS BATTERY

5.6 REPLACING THE 9V BATTERY

Figure 5-5. Replacing 9V battery

6.0 ALERTS, ALARMS, AND TROUBLESHOOTING TABLE

Yellow LED
1
ON
0
OFF
Audio Alarm
1 beep 200ms only or repeats every
20 seconds
OFF
Table 6-1. Alarm Levels
LCD
Description
Attention is
Displays Alarm Signal
NC (No Change)
MN255-1 C | 29
required
All OK

Hide quick links:

Advertisement

Table of Contents
loading