Cryopump Instruction Manual
®
 
 
 
・The warm-up time required for the 15K cryopanel to reach room temperature
depends on the kind and quantity of the gas captured in the cryopump, or the
warm-up method. 
・If a large amount of the gas is amassed by the cryopump, the inner pressure of the
cryopump may become higher than atmospheric pressure during warm-up
process in the regeneration because most captured gas evaporates for a short
time.
In this case, the pressure relief valve will open according to the following condition
to release the gas. The pressure at which the pressure relief valve begins to open
is 10 kPaG, and full open is 20 kPaG.
†For Your Information† 