5 - 2
Capacitor charging circuit
NOTICE
Charging frequency
•
Connect the oscilloscope to D 950 at TP "FR1" and "GND".
CH1
•
After switch on the following oscillogram can be measured:
Fig. 2
The capacitor charging frequency is 25 kHz and cannot be adjusted.
POLYMOBIL Plus
Signals are only present during charging of the capacitors.
TP "FR1"
5.0 V/Div
Ground at "GND"
20 µs/Div
SPR8-125.061.02
Rev. 01
Page 2 of 20
01.00
TD SD 24
Measurements
Siemens AG
Medical Engineering
5
5