Maintenance
5.1

Changing the water in the coupling circuit

Removing air pockets in the coupling bellows
PMA
Shock wave head
The water in the coupling circuit is to be changed:
Move switch S2 on board W12 to the "Empty coupling bellows" position (Y3 opens), see
the start-up instructions.
Connect the service hose with a funnel to the water tank and insert the hose end into an
available container. The water tank is emptied.
If the system parts are very dirty, clean them prior to refilling. (If necessary, remove the
tank, fill it halfway with water, and shake it to remove the dirt).
After emptying the water tank, fill the coupling circuit in accordance with the start-up in-
structions.
LITHOSTAR Multiline
RXL2-120.101.01.06.02
08.05
Maintenance
Page 22 of 52