DRYPRO MODEL 832 Test Print Check List
Date and time of
implementation
Name of a Person
in Charge
Item-
Category
Num-
ber
(1)
"Geometric pattern" is printed
(2)
Alphabetic/numeric characters on the film image can be
properly identified
(3)
No "speckle noise" was created on all or part of the film
image
(4)
No "horizontal stripe" was created on all or part of the film
Image
image
appear-
(5)
No "vertical stripe" was created on all or part of the film
ance
image
(6)
No "serious image distortion" was created on all or part of
the film image
(7)
No "image wear" was created on all or part of the film image
(8)
No "image anomaly" other than items in c to g occurred on
all or part of the film image
(9)
The "horizontal thin line" of QC pattern appears to be
resolved
Resolu-
(Use of a magnifying glass)
tion
(10) The "vertical thin line" of QC pattern appears to be resolved
(Use of a magnifying glass)
(11) After printing the QC pattern, "Dens. Error <Tray No.>" is
not
displayed on the DRYPRO 832 operation panel
(12) After printing the QC pattern, "NG" is not displayed in "1"
Density
to "4" of the "QC results" in the QC Results View on the
DRYPRO 832 operation panel, but the correct density value
(three digits) is
displayed
If the first total judgment shows "NG", print the QC pattern again after performing calibration and recheck. If the
total judgment is "NG" again on rechecking, contact Konica Minolta technical representatives.
* Enter "OK" or "NG" in the "Judgment" column and density value in "(
Year
Month Day (
) :
Item to be Checked
79
Test Print Check List
Main Unit
Name
Tray
Tray 1/Tray 2
Judgment
1st
1:(
)
2:(
)
3:(
)
4:(
)
Total judgment
1
)"
2
*
3
Recheck
4
5
7
8
1:(
)
2:(
)
9
3:(
)
4:(
)
1 0
1 1
1 2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents